4.4 Low Temperature Alarm Mode Switch: This switch gives the
operator the ability to defeat both the low temperature audio
alarm, and output connector relay operation. This feature is
useful to avoid sounding an alarm during startup or cooling
down procedures. The undertemperature LED is unaffected
by this switch. All overtemperature alarm indications are also
unaffected by this switch.
a. Low Temp. Alarm Active : In this position, undertem-
perature alarm indications operate normally.
b. Low Temp. Alarm Inhibit : In this position, an under-
temperature condition of 30 deg F (17 deg C) or more
below setpoint will
not
sound the audio alarm. Also, the
output connector relay will
not
be energized. The under-
temperature LED is unaffected, and will flash as usual.
4.5 Remote Standby Option: It is possible to wire the main frame
for remote standby operation as outlined in the following
procedure.
a. Locate the appropriate communications connector for the
zone to be occupied by the SAM. This is on the upper strip
of connectors on the rear of the model “MFC” main frame.
b. Carefully wire an isolated set of normally open relay
contacts, rated at 5 volts and1/2 amp, to pins 8 and 12
of the main frame connector. Note that this relay is
not included.
c. When the remote relay is energized, SAM will be placed in
the standby mode of operation.
d. A SAM must be installed for this option to work.